Releases
3.0.4
Version 3.0, 11 February 2021
First release; started with v3.0 to avoid confusion with older Intan software.
Version 3.0.1, 12 February 2021
Fixed bug that caused crash during electrode impedance measurement.
Version 3.0.2, 24 March 2021
Improvements to Spike Scope:
Added RMS display, consistent with old software.
Added spike rate display.
Number of spikes in recent spike history can now be as high as 500.
Each channel's recent spike history is persistent when channels are changed.
Added 'Take Snapshot' and 'Clear Snapshot' buttons to permit freezing spike history in background of display on a per-channel basis.
Added longer post-trigger spans to PSTH tool: 5s, 10s, and 20s.
Added ability to use analog inputs as digital triggers in PSTH tool.
Added ability to use analog inputs as digital triggers in Display Trigger (in Trigger tab).
Added ability to display thresholded analog inputs as digital signals in Spectrogram Tool.
In data file playback mode, added 'run immediately' option after 'jump to position'.
Extended range of y-axis spacing between waveforms on waveform display.
Fixed bug where 'Stop' button was not working during rewinding.
Fixed bug where the 'Automatically Save Trigger Channel' option in triggered recording was not working.
Fixed issue where software crashed upon exiting on some Macs.
Fixed crashing upon software startup with some smaller screens.
Improvements to TCP output:
Added ability to disable and enable individual channels for TCP output while system is running.
Improved TCP parser so that path names and file names may include spaces.
Fixed TCP output timestamps to match timestamps in recorded data.
Fixed issue where TCP data output could fill up software 'Waveform FIFO' buffer under certain conditions.
Changed TCP parameter name 'DisplayTriggerAnalogVoltageThresholdVolts' to 'TriggerAnalogVoltageThresholdVolts'.
Switched to latest version (v5.2.3) of Opal Kelly DLL.
Changes to support Linux compilation.
Windows installer:
Fixed issue where certain anti-virus programs would interfere with Windows installation.
Added additonal DLLs to Windows installation:
api-ms-win-core-winrt-string-l1-1-0.dll
api-ms-win-core-winrt-l1-1-0.dll
OpenCL.dll
Version 3.0.3, 15 April 2021
Added 50 ms option to PSTH pre-trigger and post-trigger spans.
In "One File Per Signal Type" data file format, added option to save auxiliary inputs (e.g., acccelerometer data) in amplifier.dat file.
Fixed bug that sometimes causes crashing when PC audio was enabled.
Fixed bug that did not display auxiliary inputs or supply voltages from second headstage when dual headstage adapter was used.
Expanded OpenCL compatibility to fix problems with OpenCL initialization on some systems.
Version 3.0.4, 5 August 2021
Fixed small memory leak that occurred when saving in the File Per Signal Type file format.
Improved XML settings file parsing so that files that don't include any headstage channels can still be correctly loaded.
Minor improvement to make error logging more detailed and efficient.
Modify episodic triggered recording algorithm to ignore brief trigger-off events.
Improved optimum SPI delay finding algorithm so that setting delay manually is less likely to be necessary.
In data file playback mode, electrode impedances are now displayed.
Fixed issue with external fast settling in RHD systems not properly triggering as a result of digital input activity