CONTENTS
--------
1 Resolved Issues
2 On-board Detector Server Compatibility
3 Firmware Requirements
4 Kernel Requirements
5 Download, Documentation & Support
1 Resolved Issues
Firmware
--------
* [Jungfrau] Stopping acquisition in sync mode
Previously, stopping acquisition in sync mode would result in the slave
status remaining in Running status for next acquisition and framesl (frames
left) would be stuck from the previous acquisition. This is fixed in firmware.
Detector Server
---------------
* [Jungfrau] Changing master to slave in sync mode
When in sync mode, changing from master to slave would change the status
to Running in firmware. A workaround in the detector server fixes this.
* [Jungfrau] Row and column for multi module Jungfrau
This mainly affected only the GUI, unless one was using the row and column
sent out by the detector. Row and column values were mixed up, but fixed now.
* [Jungfrau] Timing in sync mode
When sync is enabled, setting or getting timing would give an error. This
is fixed now.
Client
------
* [Jungfrau] Multi module sync mode
For start, stop and trigger acquisition, the procedure to start the slaves
(or the master for stop acquisition) first had to be obeyed. This is fixed
now.
ZMQ
---
* [Jungfrau][Gotthard2] Client ZMQ port when using 2 interfaces
When switching between 1 and 2 interfaces, the client or gui zmq port
was not set properly and resulted in dummy streaming forever. This is
fixed now.
* Rx_roi
This parameter is now also streamed in the zmq header to be able to use in
external GUIs.
GUI
---
* [Jungfrau] Show Fix_G0 in expert mode
Fix_G0 was only shown when expert mode was enabled from Settings Tab. Now,
fixed to show when expert mode enabled from any tab.
2 On-board Detector Server Compatibility
Eiger 7.0.0
Jungfrau 7.0.2
Mythen3 7.0.0
Gotthard2 7.0.0
Gotthard 7.0.0
Moench 7.0.0
Ctb 7.0.0
On-board Detector Server Upgrade
--------------------------------
From v6.1.0 (without tftp):
Using command 'updatedetectorserver'
From 5.0.0 (with tftp):
Using command 'copydetectorserver'
Instructions available at
https://slsdetectorgroup.github.io/devdoc/serverupgrade.html
3 Firmware Requirements
Eiger 20.02.2023 (v31)
Jungfrau 16.05.2023 (v1.4.1, HW v1.0)
15.05.2023 (v2.4.1, HW v2.0)
Mythen3 24.01.2023 (v1.4)
Gotthard2 23.11.2022 (v0.3)
Gotthard 08.02.2018 (50um and 25um Master)
09.02.2018 (25 um Slave)
Moench 05.12.2022 (v0.3)
Ctb 05.12.2022 (v1.1)
Detector Upgrade
----------------
The following can be upgraded remotely:
Eiger via bit files
Jungfrau via command <.pof>
Mythen3 via command <.rbf>
Gotthard2 via command <.rbf>
Moench via command <.pof>
Ctb via command <.pof>
Gotthard cannot be upgraded remotely
Except Eiger,
upgrade
Using command 'programfpga' or
udpate both server and firmware simultaneously
Using command 'update'
Instructions available at
https://slsdetectorgroup.github.io/devdoc/firmware.html
4 Kernel Requirements
Blackfin
--------
Latest version: Fri Oct 29 00:00:00 2021
Older ones will work, but might have issues with programming firmware via
the package.
Nios
-----
Compatible version: Mon May 10 18:00:21 CEST 2021
Kernel Upgrade
---------------
Eiger via bit files
Others via command
Commands: udpatekernel, kernelversion
Instructions available at
https://slsdetectorgroup.github.io/devdoc/commandline.html
https://slsdetectorgroup.github.io/devdoc/detector.html
https://slsdetectorgroup.github.io/devdoc/pydetector.html