Brew and Calibrate iSpindel
If this feature is enabled, BPL will record the TILT angles from iSpindel, expect the gravity data from user, and derive formula to calculate the gravity automatically. The procedure to use "Brew and Calibrate" is
- Setup iSpindel to make it report to BPL.
- Get the Tilt value in water(SG:1.0) before dropping it into the fermenter.
- Measure the Original Gravity.
- Enable "Calibrate iSpindel" and enter the Tilt value in water at the Gravity Device pages. This must be done before startting logging.
- Start local logging.
- Input SG, which is the value of Original Gravity. Note: You should usually input the same value as OG. The required data is SG here.
- Measure SG as usual or more frequently. Input the measured SG.
BPL will derive the formula by the Tilt values and gravity data input. If the number of data pairs is less than 3, then linear(x) formula is used. Second order polynomial(x^2) will be derived if the number of pairs is equal to 3. If 4 or more gravity readings are available, 3 order polynomial will be derived. At the beginning, there should be two readings, 1.0 and OG. Even though you can't expect precise and correct gravity readings from the chart from the beginning, the change of gravity can be deduced by the change of Tilt values.
The gravity reading to input shoud be NON temperature-corrected. The wort used to measure gravity shoud be at the same temperature as the fermenting wort. The data used to calibrate iSpindel is from the iSpindle in the fermenting wort and the gravity input. The derived formula will be used to calculate the gravity value. Temperature correction is applied on the calculated values.
Note: in Brew and Calibrate mode, the gravity values displayed on the chart is calculated by the browser, or by Javascript to be exact. That gravity data can't be used in Beer Profile. The Beer Profile will use the data that users input.
If BPL is in calibrating mode, there will be a lower case "f" at the bottom of the legend block. Click the "f" will bring out the calibration points and formula used.
Wiki is deprecated. Check doc directory for latest document. https://github.com/vitotai/BrewPiLess/tree/master/doc
- Hardware Setup
- Localization of web frontend
- Flash layout and New Frontend
- Software Installation
- WiFi Setup
- System Setup
- Device Setup
- Using Buttons & Menu
- Temperature Logging
- Cloud Logging
- Beer Profile
- iSpindel Support
- Manual Gravity Logging
- Use with Glycol
- Auto Capping(Spunding)
- MQTT Remote Control
- Pressure Monitor and Control
-
Other URLs
- Clear WiFi setting
- Format file system
- OTA update
- Overwrite pages
-
JSON Commands
- Temperature Unit
- Sensor Calibration
- FAQ