Last revised: 20180619 (K Wong)
Perform instructions in the correct order!!
- Connect
OXY10
to a power source - Close all applications on the computer (PC ONLY)
- Connect
OXY10
to the computer - Launch the
PreSens Measurement Studio 2
Program on the computer - Remove sensor covers and connect to
OXY10
- Oxygen sensor installation
- Line up notch on the side of the cable to the opening in the port
- Push in the metal part on the cord and guide the notch through the grooves
- Temperature sensor installation
- Line up red dots on the cable and the port
- Make sure to slide into the port straight
- Check side of screen for device name text indicating the device is connected:
OXY-10 ST, 02
SABD0000000001
- Click on device name and it will open the following screen
Each Channel will run an individual sensor. Each Channel requires the following 3 things to be set:
- Measurement
- Sensor
- User
- Click
Measurements
on the top tool bar - Then click
New
- Assign a new name and click
Save
- Make sure the new Measurement is highlighted and click the desired channel you want to pair with
- Click
Assign
- Your new
Measurement
should be paired with the desired channel
- Locate the correct calibration information to the specific sensor
- Click
Sensors
on the tool bar - Then click
New
- Fill out the correct calibration information as on the sheet
- Sensor Name: OXY_CHX, where x = channel name
- Sensor type: Pst7-10
- Batch ID: Found on the barcode and under Batch number at the top of the calibration sheet
- Lot Nr: Leave this empty
- Calibration Data
- pATM: Atmospheric Pressure from calibration sheet
- pATM Unit: hPA
- Mode: Humid
- Cal0
- T0
- Cal2nd
- T2nd
- O2-Cal2nd: Add the value of oxygen solubility for the temperature and salinity of the water you are using (e.g., 203.2 µmol L-1 for 36 ppt at 27°C) - Measure seawater temperature and salinity prior to setting this and input those values.
- Oxygen unit: μmol/L
- Sensor Constants
- f1: fill with value on calibration sheet
- dPhi1: fill with value on calibration sheet
- dKSV1: fill with value on calibration sheet
- m: fill with value on calibration sheet
- dPhi2: fill with value on calibration sheet
- dKSV2: fill with value on calibration sheet
- Click
Save
- Make sure the new sensor is highlighted and click the desired channel you want to pair with
- Click
Assign
- Your new
Sensor
should be paired with the desired channel
- Click
Users
on the tool bar - Choose default user with default parameters
- Ensure the channel has a measurement, sensor, and user (default)
- Click on the channel and click
Live View
- Make sure to change:
- Units to μmol/L
- Set Salinity to the value of the seawater you are measuring (e.g. 36%)
- The click
Start